MEASURE USING CONTINUOUS SWEEP
Automatic recording of measurement values during continuous sweeping of the shaft.
There is no limit on the number of points.
Preparations
Before you start measuring, make sure you have done the preparations you need.
l
Mount the measuring units. "Continuous sweep" is available when you use XT60 measuring units.
l
To calculate results, you need to enter at least the distance between the measuring units, see "Enter distances"
on page 36.
l
Connect measurement units.
l
If needed, perform a rough alignment. See "Set up measuring units" on page 34.
l
If needed, measure Softfoot. See "Softfoot" on page 37.
Measure
It is possible to switch measuring method before you have registered a value.
1. Tap
on the tab to select the method Continuous sweep.
2. Adjust laser to the center of the targets. If needed, adjust the units on the rods, then use laser adjustments knobs.
3. Tap
to start the measurement.
4. Turn the shafts. Turn the shafts as much as possible for a more accurate result.
5. Tap
to stop the measurement.
6. Tap
to go to the Result view, or tap
to measure again.
The registered values are saved when you leave the Measure view. If you return to the Measure view, it is possible to
do a new measurement.
A. Tap to display detector information.
B. Yellow area is where points have been registered.
C. Measurement direction. If you change direction during the measurement, the arrow turns red.
D. Quality assessment. See "Quality assessment" on page 52
E. Tap to display detailed information.
F. Tap to start and stop the measurement.
Make a new measurement. This makes it possible to check the repeatability of the measurement.
Measure using EasyTurn™.
47
HORIZONTAL
Summary of Contents for XT11
Page 1: ...User Manual English 05 0835 EN Revision 2 0 User Manual last built 9 12 2017...
Page 2: ......
Page 6: ...This page intentionally left blank to ensure new chapters start on right odd number pages...
Page 32: ...32...
Page 58: ...A Offset B Angular C Both offset and angular misalignment 58 BELT...
Page 74: ...Part No 12 0985 Camera removed from XT11 74 TECHNICAL DATA...