MEASURE USING EASYTURN™
With EasyTurn™, it is possible to measure with as little as 40º spread between the measurement points. However, for
an even more accurate result, try to spread the points as much as possible.
Preparations
Before you start measuring, make sure you have done the preparations you need.
l
Mount the measuring units.
l
To calculate results, you need to enter at least the distance between the measuring units, see "Enter distances"
on page 36.
l
Connect measurement units.
l
If needed, perform a rough alignment. See "Set up measuring units" on page 34.
l
If needed, measure Softfoot. See "Softfoot" on page 37.
Measure
It is possible to switch measuring method before you have registered a value.
1. Tap
on the tab to select the method EasyTurn.
2. Adjust laser to the center of the targets. If needed, adjust the units on the rods, then use laser adjustments knobs.
3. Tap
to register the first position. A red marking is displayed.
4. Turn the shafts at least 20º.
5. Tap
to register the second position.
6. Turn the shafts at least 20º.
7. Tap
to register the third position.
8. Tap
to go to the Result view, or tap
to measure again.
The registered values are saved when you leave the Measure view. If you return to the Measure view, it is possible to
do a new measurement.
A. The Measure icon is active in the workflow.
B. Tap to display detector information.
C. Red = turn shafts outside the red marking.
Green = turn shafts to green area.
Blue = registered position.
D. Measurement table.
41
HORIZONTAL
Summary of Contents for XT11
Page 1: ...User Manual English 05 0835 EN Revision 2 0 User Manual last built 9 12 2017...
Page 2: ......
Page 6: ...This page intentionally left blank to ensure new chapters start on right odd number pages...
Page 32: ...32...
Page 58: ...A Offset B Angular C Both offset and angular misalignment 58 BELT...
Page 74: ...Part No 12 0985 Camera removed from XT11 74 TECHNICAL DATA...