SERVICE MANUAL
2–8
July 1995 – 1C7833
Replacement of the DIVERTERS or the BEARINGS
Note
For replacement of the SOLENOIDS, see Page 2–9.
[1] De-energize the PROCESSOR. See Page 2–5.
[2] Do the General Access procedure on Page 2–6 as required.
Possible damage from electrostatic discharge.
ESD
[3] Disconnect the CONNECTOR for the SOLENOID from the 100 BOARD.
Caution
Do not allow the SCREW or the SPRING to fall. You may want to place a cloth over the processing section to catch
falling parts
[4] Remove:
(a) SPRING and SCREW
(b) SOLENOID with the SOLENOID BRACKET and the DIVERTER
(c) DIVERTER from the SOLENOID BRACKET
(d) front BEARING from the SOLENOID BRACKET
(e) back BEARING from the SORTER FRAME
Figure 2–6
Replacement of a DIVERTER
[5] Reverse the above steps to install a new DIVERTER and BEARINGS. While installing, do the following:
(a) Check that the SOLENOID BRACKET is inserted in the long hole in the SORTER FRAME.
(b) Check that both ends of the SPRING are in the grooves.
(c) When tightening the SCREW, hold the SOLENOID BRACKET so that the notch is fully seated around
the locating TAB. See Figure 2–6.
H127_2023HA
groove
groove
SORTER FRAME
notch
SCREW
TAB
DIVERTER
back
BEARING
front
BEARING
hole
long
CONNECTOR
SPRING
BRACKET
SOLENOID
SOLENOID