SERVICE MANUAL
1–18
July 1995 – 1C7833
Figure 1–21
Installation of the New COVER on the ELECTRICAL BOX
[2] Place the ELECTRICAL BOX on a table or work surface.
Possible damage from electrostatic discharge.
ESD
[3] Insert the WRIST STRAP into the ESD GND JACK on the ELECTRICAL BOX, and place the WRIST STRAP
around your wrist. See Figure 1–19 on Page 1–17.
[4] Loosen the CAPTIVE SCREW and open the COVER.
[5] Remove:
• the SCREWS holding the 500 BOARD to the COVER
• the SCREWS that hold the STRAIN RELIEFS
• the 2 NUTS and 2 GROUND STRAPS from the COVER
[6] Remove the 3 SCREWS from the front of the existing COVER of the ELECTRICAL BOX.
[7] Lift and hold the 500 BOARD.
[8] Remove the existing COVER, and place the new COVER in position under the 500 BOARD.
Note
The existing COVER will not be used again.
[9] To fasten the new COVER to the ELECTRICAL BOX, install:
• 3 SCREWS
• 2 GROUNDS STRAPS and 2 NUTS on the new COVER
• 500 BOARD and the STRAIN RELIEFS on the COVER
[10] Check that the COVER is in the correct position by closing the COVER and tightening the CAPTIVE SCREW.
H127_2021HA
new
COVER
STRAIN
RELIEF
2 GROUND
STRAPS
3 SCREWS
CONNECTOR
ELECTRICAL BOX
HARNESS
STRAIN RELIEF
NUT
NUT
SCREW
CAPTIVE SCREW
H127_2021HCA