INSTALLATION INSTRUCTIONS
34
February 1996 – 1C7836
Note
If the RECIRCULATION PUMP does not energize, check that the FIXER and DEVELOPER TANKS are full. If the
TANKS are full and the RECIRCULATION PUMP does not energize, do the following steps:
a. Connect the PORTABLE COMPUTER, using the INTERFACE CABLE TL-4391 and the DIAGNOSTICS
DISKETTE 5B6278. Enter the “Specific Mode” to manually operate the RECIRCULATION PUMP.
b. To allow correct operation of the LEVEL SENSOR PROBES, to clear air from the plumbing system, and to mix
the solutions, energize the RECIRCULATION PUMP and wait approximately 5 seconds.
c. Return the PROCESSOR to normal operating mode.
[8] With the EXTERNAL COVERS removed, check the complete plumbing system for leakage.
Note
For instructions on connecting the PORTABLE COMPUTER and using the DIAGNOSTICS DISKETTE 5B6278, see
the User Instructions packed with the DIAGNOSTICS DISKETTE.
[9] Enter the “Specific Test Mode” and select the “Processor Component Test”.
[10] Check the operation of:
• REPLENISHMENT PUMPS
• RECIRCULATION PUMPS
• DRYER HEATER and BLOWER
• WASH INPUT WATER SOLENOID
[11] Enter the “Processor Sensor Test Mode”.
[12] Check the operation of:
• FILM DETECTOR SENSORS 1 and 2
• EXIT SENSOR
• COVER SWITCH
• DEVELOPER and FIXER TANK SENSORS
[13] If installing a 180 LP PROCESSOR, advance to Step
. If the PROCESSOR is a 180 LPS
PROCESSOR, check the operation of the SORTER by doing the next procedure.
H127_0077BA
H127_0077BCA
INTERFACE CABLE
ELECTRICAL BOX