Checking for Correct Operation
1C7836 – February 1996
33
Section 4: Checking for Correct Operation
Checking the Components and SENSORS
[1] Close the DEVELOPER and FIXER DRAIN VALVES.
[2] To allow correct operation of the LEVEL SENSORS, add 470 mL (16 fl oz) of developer to the DEVELOPER
TANK and 470 mL (16 fl oz) of fixer to the FIXER TANK.
[3] Slowly, fill the PROCESSOR DEVELOPER and FIXER TANKS with clean water to the overflow limit.
[4] Move the wall POWER SWITCH to the “ON” position.
[5] Move the MAIN CIRCUIT BREAKER CB1 on the PROCESSOR to the “|” position.
Important
The PROCESSOR initializes a self-check procedure.
[6] Check that the WASH WATER SOLENOID, DRYER BLOWER, DRYER HEATER, and RECIRCULATION
PUMP energize.
[7] Check that solutions can drain smoothly from all drain and OVERFLOWS.
H127_0076BA
H127_0076BCA
FIXER DRAIN VALVE
DEVELOPER DRAIN
VALVE