Maximizing the Accuracy of the Film Logging Feature
Although up to 3 of the SENSORS on the 6000 BOARD may malfunction before a “Film Accumulator LED Error” will
appear on the DISPLAY PANEL, all of the SENSORS must be operating correctly to ensure the maximum accuracy
of the Film Logging feature.
To check that all the SENSORS are operating correctly, use the PROCESSOR’S internal diagnostics. See the
Diagnostics Manual, Publication Number 5B6333. When you do the FILM ACCUMULATOR test, any of the following
status indicators may appear on the DISPLAY PANEL of the PROCESSOR:
+ indicates that film is detected
- indicates that film is not detected
? indicates that the LED SENSOR is not operating correctly
If one of the SENSORS is not operating correctly and you require optimum accuracy of the Film Logging feature, you
will need to install a new 6000 BOARD.
Important
• To ensure optimum accuracy of the Film Logging feature, align films with either the left or the right edge of the
FILM GUIDE when you feed them into the PROCESSOR.
• Because the metric and English films are often very close in size, the PROCESSOR cannot always differentiate
between the following film sizes:
– 24 x 30 cm and 10 x 12 in.
– 30 x 35 cm and 11 x 14 in.