ADJUSTMENTS AND REPLACEMENTS
74
01JUN99 – 5B6331
Bypassing or Resetting the Access Code
Access Code
In order to enter the PROCESSOR’S internal diagnostic procedures, you must either enter the correct access code,
or know how to bypass or reset the access code in case the operator has changed the access code from the default
of 4-2-3-1. This procedure outlines both how to bypass and how to reset an access code.
The Service Bypass procedure allows you to access all processor functions without knowing or changing the access
code set by the operator.
The User Bypass procedure allows you to reset the access code to the default of 4-2-1-3 if the operator forgets the
new access code he or she has entered.
Service Bypass
Warning
Dangerous Voltage
[1] Energize the PROCESSOR.
[2] Move the first SWITCH on SWITCH S5002 on the 5000 BOARD to the alternate position.
Note
The original position of the first SWITCH is not important. The software only recognizes a change in the position of
the SWITCH.
User Reset
Warning
Dangerous Voltage
[1] Energize the PROCESSOR.
[2] Move the fourth SWITCH on SWITCH S5002 on the 5000 BOARD to the alternate position.
Note
The original position of the fourth SWITCH is not important. The software only recognizes a change in the position
of the SWITCH.
+
-
1 2 3 4
H150_0061BA
H150_0061BCA
S5002
SWITCH