Users manual
ET1260B and ET1260A
6 1/2 Digit Multimeter
32
off power, the upper limit value will be set as 1 and lower limit value as 0.
In configuration limit test, press
(Limit) to enter setting interface, and edit values
through direction keys.
Press the softkey of Upper limit (
) to set upper limit value.
Press the softkey of Lower limit (
) to set lower limit value.
Press softkye of On/Off (
/
) to start or stop limit operation. After starting, limit test status
indication can be seen in the right side of main interface, with
Pass
meaning that the measured
value does not exceed the specified upper or lower limit, with
Excessively higher
meaning
exceed the specified upper limit and
Excessively lower
meaning exceed the specified lower limit.
When status switches from
Pass
to
Excessively higher
or
Excessively lower
, the beeper will
beep one time.
After setting, press the softkey of Exit (
) to go back to the former-layer interface.
3.3 Trigger
ET1260B and ET1260A provide many trigger methods, including auto, single and external.
Multimeter can read one reading or readings in specified number (up to 50,000) every time it
receives a trigger signal, and can set the delay time between trigger and readings.
In measurement interface, press
to enter the trigger configuration interface indicated in
the following figure.
Press the softkey of Trigger source (
) to set auto, single or external trigger.