Users manual
ET1260B and ET1260A
6 1/2 Digit Multimeter
23
3. Re-connect the probe end of positive lead to test circuit, and measure capacitance value.
2.1.9 Continuity measurement
Basic information
Test power source: 1mA.
Beeper threshold: 1Ω-1000Ω is available, and 10Ω as a default.
Operation description
1) Press
to select continuity measurement function.
Description: When the measured resistance value is larger than the beeper threshold, display
OPEN
, and when the measured resistance value is smaller than the beeper threshold, display
resistance value.
2) Connect test lines as indicated in the chapter of “1.4 Measurement wiring”.
3) Press the softkey of Setup (
) to set beeper threshold: when the measured resistance is
lower than beeper threshold, the beeper beeps, and the beeper threshold can be set as 1Ω-1000Ω,
with 10Ω being a default.
2.1.10 Diode measurement
Basic information
Test current source: 1mA.
Beeper threshold: 0.3V
≦
test voltage
≦
3V, non-adjustable.
Operation description
1. Press
to select diode measurement function.
Description: When voltage value is higher than measurement threshold, display
OPEN
,
otherwise display voltage value.
2. Connect test lines as indicated in the chapter of
1.4 Measurement wiring
.