7) If necessary, press
to configure math operation function (limits and relative), and refer
to description in the chapter of
Math operation
for details.
Leads resistance eliminating method for two-wire resistance measurement:
1. Connect one end of test lead to multimeter, and short together with probe end.
2. Press the softkey of Relative (
) to make zero, and every resistance measurement will
reduce current lead resistance measurement value.
3. Connect test probe end to test circuit, and measure resistance value.
2.1.7 Frequency/Period measurement
Basic information
Frequency measurement band: 3Hz to 300 kHz.
Period measurement band: 3.3us to 0.33s
Measurement range: 100mV, 1V, 10V, 100V and 750V
Input signal range: 100mVAC to 750VAC.
Technology: equal precision.
Operation description:
1) Press
to select frequency measurement function or period measurement function.
Description: During frequency or period measurement, auxiliary display refers to virtual value of
AC voltage of input signal.
2) Connect test lines as indicated in the chapter of
1.4 Measurement wiring
.
3) Select measurement range as required, and may select automatic measurement, or press the
softkeys of Measurement range+ (
) and Measurement range- (
) to select fixed
measurement range, which have same effect as keys of
and
.
4) Press the softkey of Relative (
) to record the current value, after which display the