AQ COMFORT Instruction & Operations Manual
Version 1.01
Page 39 of 44
4. OPTIONAL EXTERNAL MEASURING PROBE
Certain applications might have the need for the optional probe
(12” long stainless steel probe with 10’ ft long hose) for hard to
reach places, separate areas, etc. The probe consists of the
stainless steel extraction probe, the sampling line and a Type K
thermocouple for external temperature measurements with the
probe.
This standard probe (Part # AACSF22) is suitable for temperatures up to 800ºC/1470ºF.
The end of the probe’s sampling line connects to the “A” inlet connector on the bottom
of the AQ Comfort and the thermocouple connector connects to the TcK
TEMPERATURE 1 input on the bottom of the AQ Comfort.
5. OPTIONAL L-SHAPED PITOT TUBE
Pitot tube –
This optional L-shaped pitot tube consists of stainless steel tubing with the
tips open and bent at a certain angle to properly measure air and/or gas velocities. The
Pitot tube must always be oriented with the open tips parallel to the direction of the air
flow. The end of the Pitot tube assembly is connected by means of two flexible hoses to
a very accurate pressure transducer located inside the monitor.
6.
OPTIONAL DIFFERENTRIAL PRESSURE HOSES
(part #: AACKP01)
This optional PRESSURE HOSES consist of two (2) clear hoses
for either (1) single positive pressure connection, one (1) single
negative pressure connection, or (2) hoses for differential pressure
using both connections simultaneously. The hoses connect to the
(-) and/or (+) PRESSURE connections on the Bottom of the AQ
Comfort.