AQ COMFORT Instruction & Operations Manual
Version 1.01
Page 41 of 44
CHAPTER 13
BASIC TROUBLESHOOTING
ZERO ERRORS
Channel
Possible Causes
Resolution
(Electrochemical sensors)
CO
Sensor has been recently
exposed to gas and has
not yet returned to zero.
Purge for 10 minutes with
fresh air, monitor sensor
voltage, and re-zero.
Insufficient Warm Up
Allow the sensors to warm
up after 10 minutes
Incorrect ZERO
calibration
Repeat ZERO cal in fresh air
or with ZERO AIR cylinder
Pump Failed
Check pump flow, if pump
failed replace
Battery was dead, sensor
has destabilized.
Charge battery, wait up to 24
hours for sensors to
stabilize, and re-zero.
Sensor cell life has
expired.
Call E INSTRUMENTS
replacement.
Temperature Thermocouple Thermocouple is not
connected
Check electrical connections
running to the probe
Infrared CO
2
No response from infrared
system
Zero period must be at least
45 seconds.
Zero the
instrument again with a
cylinder ZERO AIR
.
Infrared system is
reporting error code
Infrared system may need
maintenance.
Velocity
Velocity probe/Pitot Tube
is not connected
Check probe and
connections.
Bluetooth connection failure Connection Failed
Redo connection process