AQ COMFORT Instruction & Operations Manual
Version 1.01
Page 28 of 44
ZERO AIR Gas Bottle Calibration:
D. CALIBRATION PROCEDURE
The following page illustrates the sequence of key strokes to carry out a span
calibration of the AQ Comfort. It is assumed that the instrument has been zeroed and
there have been no error messages.
Electrochemical Gas Sensors
1. Turn the AQ Comfort on and press Measure. Allow the instrument to warm up for
15 minutes before beginning the calibration procedure.
2. Connect the outlet from the tee fitting of the gas calibration fixture to the “A” inlet
connector on the bottom of the monitor using the set up shown in the above
calibration apparatus photo:
NOTE: Please take special care when performing Calibration of the monitor.
You cannot revert back to the factory setting after making any adjustments!
Gas Hose from Cylinder connects to
Center “A” Inlet on Bottom of AQ
Comfort for Span Calibration
T – for
proper
air/press
ure flow