User Manual HTS801
Humidity and Temperature Sensor for High Humidity and Chemically Polluted Conditions | 21
▪.
HTTP-Webserver: Port 80
For a quick communication check, enter the desired IP address in a web browser and connect with the
HTS801 Ethernet module’s Webserver.
Fig. 23 E+E net web interface screenshots
Alternatively, send an ICMP echo request (“ping”) to check correct communication settings.
IPv4 Settings
Factory settings
IP Address
192.168.0.64
Subnet Mask
255.255.255.0
Gateway
192.168.0.1
DNS
192.168.0.1
Tab. 3
IPV4 settings
DHCP options can be set with jumper on the electronics board, refer to Fig. 24. Factory setting: DHCP disabled
(static IP).
Fig. 24 Ethernet module - DHCP setting
The setup of the Modbus TCP communication can be performed via PCS10 Product Configuration Software or via
display and push-buttons, see chapter 11.3 Optional Menus Fig. 57 IP settings.
Modifying the IP address via PCS10 or display is possible only when the DHCP jumper is set to “Static”.
Otherwise the IP settings are read-only.