background image

11

User Manual EE872 Modular Probe for CO

2

, RH, T and p

Decoding of float values:

Float values are stored according IEEE754 standard. The byte pairs 1, 2 and 3, 4 are inverted as fol-

lows:

MMMMMMMM

MMMMMMMM

SEEEEEEE

EMMMMMMM

 Byte 3

Byte 4

Byte 1

Byte 2

Example:

 

Response [Hex]

Value in decimal

 Byte 1 (Register 2 - Hi)

Byte 2 (Register 2 – Lo)

Byte 3 (Register 1 - Hi)

Byte 4 (Register 1 - Lo)

44

A1

0B

 00

1288.34375

See also Application Note AN0103 at 

www.epluse.com/EE872

.

5.3  Error Indication on the Analogue Output (NAMUR)

The EE872 features an error indication on the analogue output according to the NAMUR NE 043 

recommendations (Standardization of the Signal Level for the Failure Information of Digital Transmitters, 

Edition 2003-02-03, see 

www.namur.net/en/recommendations-and-worksheets/current-nena.html

).  

The feature is disabled by factory default can be enabled with the EE-PCS Product Configuration 

Software, see above.

Output signal

NAMUR signal level

0-5 V

5.5 V

0-10 V

11 V

4-20 mA

21 mA

0-20 mA

21 mA

6  Maintenance

6.1  Replacing the Sensing Module EE872S

If needed, the sensing module can be replaced by a new one. Please refer to chapter 2.2 for the order 

codes.

Please note:

 

The new sensing module EE872S must feature same CO

2

 measuring range as the original EE872 probe! 

If the measuring range of the replacement module and of the original EE872 probe are different, the 

analogue output will stay at 4 mA, 0 V or NAMUR error indication while the CO

2

 reading via RS485 

interface will be 0 ppm.

Procedure: 

     

»  Power off the EE872.

»  Remove the filter cap by  

  turning it counter-clockwise.

»  Remove the sensing module by

  pulling it straight out from the output unit.

»  Plug the new EE872S sensing module 

  into the output unit.

»  Screw the filter cap fingertight onto the probe.

Fig. 5 

EE872 modular construction

output unit

sensing module

filter cap

Summary of Contents for EE872 Series

Page 1: ...BA_EE872_e V1 2 technical data are subject to change User Manual EE872 Modular Probe for CO2 RH T and p...

Page 2: ...equipment has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmf...

Page 3: ...put and RS485 Interface 7 4 Installation 8 5 Setup and Adjustment 8 5 1 EE PCS Product Configuration Software 9 5 2 Digital Interface RS485 with Modbus RTU Protocol 9 5 2 1 Digital Settings 9 5 2 2 Mo...

Page 4: ...nce In any case the scope of liability is limited to the corresponding amount of the order issued to the manufacturer The manufacturer assumes no liability for damages incurred due to failure to compl...

Page 5: ...vices are not accidentally used by marking them clearly as faulty A faulty device may only be investigated and possibly repaired by qualified trained and authorized staff If the fault cannot be fixed...

Page 6: ...ng is activated the RH and T data is not available The calculated Td value is still available All values are available on the digital interface All versions provide CO2 measurement T and p compensated...

Page 7: ...n pin not connected EE872 set to RS485 interface option P1 in the order code features RS485 interface EE872 set to analogue outputs option GA7 or GA11 in the order code the RS485 interface is active f...

Page 8: ...with mounting clip HA010227 EE872 with mounting flange HA010226 For outdoor applications the EE872 must be used with the radiation shield HA010510 not included in the scope of supply see data sheet A...

Page 9: ...Device address baud rate parity and stop bits can be set via 1 EE PCS Product Configuration Software and the Modbus configuration adapter HA011018 The EE PCS can be downloaded free of charge from www...

Page 10: ...FA1 Temperature T F 50 4003 0xFA2 Temperature T K 50 4005 0xFA4 Dew point temperature Td C 100 4053 0xFD4 Dew point temperature Td F 100 4054 0xFD5 Dew point temperature Td K 100 4074 0xFE9 Tab 3 Inte...

Page 11: ...abled by factory default can be enabled with the EE PCS Product Configuration Software see above Output signal NAMUR signal level 0 5 V 5 5 V 0 10 V 11 V 4 20 mA 21 mA 0 20 mA 21 mA 6 Maintenance 6 1...

Page 12: ...rotection cap rubber caps Fig 6 Calibration adapter HA010785 with mounted rubber caps for use as protection cap Procedure Remove the filter cap and install the calibration adapter onto the probe Conne...

Page 13: ...range 40 60 C 40 140 F Accuracy3 5 60 C 41 140 F typ 0 5 C 0 9 F Outputs Analogue CO2 only 0 5 V 0 10 V 1 mA Load current 1 mA 0 20 mA 4 20 mA 3 wire Load resistance 500 Ohm Digital interface CO2 RH T...

Page 14: ...0 6460 Office Stuttgart Tel 49 151 538 37 500 E E Elektronik Italy info epluse it Tel 39 02 2707 86 36 E E Elektronik France info epluse fr Tel 33 4 74 72 35 82 E E Elektronik USA office epluse com Of...

Reviews: