27
Er39
Error detected during EAROM write operation. This error may appear in operating mode when
storing new value in EAROM (for example, alarm threshold or transducer calibration). The new
values will be enabled but they will be lost when the instrument is powered down. This error
message disappears automatically after 10 seconds. If the error appears during
configuration/calibration, press
FUNC
or
RESET
to restart the procedure and then repeat
operations. If the error persists, return the instrument to your supplier.
Er40
This error occurs during an incomplete configuration save operation. The instrument had its
power reset during a non-volatile ram write cycle. To clear, initiate the default data loading
procedure and repeat the configuration and calibration procedure.
Er101
The configuration data stored in EAROM is wrong or inconsistent. This error may appear at
instrument start-up in operating mode. This error message disappears automatically after 3
seconds. After that the instrument will reset.
If the error persists, enable configuration/calibration with the internal switch, load the default
configuration data, and then perform a new configuration.
Er312
Error during internal autozero measurement for temperature drift compensation. The instrument
repeats this check every 3 seconds. The analog retransmission and alarm go low scale or high
scale as a fail safe configuration. If the error persists, return the instrument to Dynisco.
Er313
Calibration data fails checksum. To correct, initiate the default data loading procedure and
recalibrate.
Er314
Instrument non-volatile memory un-initialized. Return to Dynisco.
Er315
Instrument non-volatile memory un-initialized. Return to Dynisco.
Er316
Instrument non-volatile memory un-initialized. Return to Dynisco.