Page 68 of 146
198-1600-00 Rev F
values for any calibration point can be viewed at any time without changing the values
(select “no” when prompted to store changes).
6.
After completing a calibration the calibration values should be saved by selecting “yES”
when prompted to store changes. The calibration values are stored to permanent memory
and will be retained through power on and power off cycles.
Cal
Point
Calibration Menu Display
Calibration point(s) Description
“rd A” or “rd b” Calibration (Read
back meters and output)
CL01
“I.rd OFF- -SET CL01”
Module A: Current reading offset
CL02
“E.rd OFF- -SET CL02”
Module A: Voltage reading offset
CL03
“I.rd SEt CL03”
Module A: Current setting used for current
gain calibration
CL03
“I.rd gAin CL03”
Module A: Current gain value
CL04
“E.rd SEt CL04”
Module A: Voltage setting used for voltage
gain calibration
CL04
“E.rd gAin CL04”
Module A: Voltage gain value
FLtA
“rEAd FILt SEC”
Module A: Readings filter size in seconds,
smaller value provides faster updates, larger
value provides more filtering.
CL05
“I.rd OFF- -SET CL05”
Module B: Current reading offset
CL06
“E.rd OFF- -SET CL06”
Module B: Voltage reading offset
CL07
“I.rd SEt CL07”
Module B: Current setting used for current
gain calibration
CL07
“I.rd gAin CL07”
Module B: Current gain value
CL08
“E.rd SEt CL08”
Module B: Voltage setting used for voltage
gain calibration
CL08
“E.rd gAin CL08”
Module B: Voltage gain value
FLtA
“rEAd FILt SEC”
Module B: Readings filter size in seconds,
smaller value provides faster updates, larger
value provides more filtering.
“OPt” Calibration (Analog Option
Card)
CL09
“I.Hi CtL CL09”
Upper analog current control input (optional)
CL10
“I.Hi rEAd CL10”
Upper analog current reading output
(optional)
CL11
“I.Lo CtL CL11”
Lower analog current control input (optional)
CL12
“I.Lo rEAd CL12”
Lower analog current reading output
(optional)
CL13
“E.Hi CtL CL13”
Upper analog voltage control input (optional)
CL14
“E.Hi rEAd CL14”
Upper analog voltage reading output
(optional)
CL15
“E.Lo CtL CL15”
Lower analog voltage control input (optional)
CL16
“E.Lo rEAd CL16”
Lower analog voltage reading output
(optional)