24
RMO100A Error Messages
8.4 Error Message "Change Current"
This message is displayed if during the test the voltage at the test object exceeds 5 V. In this case
reduce the test current and repeat the test.
8.5 Error Message "Overheat"
This message is displayed when RMO100A
‟s operating temperature rises too high due to a high
ambient temperature or, despite the duty cycle time control, too many high current tests were
performed in too short of a time.
In this case, wait for the test set to cool down and repeat the test.
8.6 Error Message "Malfunction"
If this message is shown on the display, or if RMO100A cannot be operated anymore at all, a
serious internal error occurred.
Please do not open RMO100A by yourself. Contact the IBEKO Power AB (refer to section
”Manufacturer Contact Information”).
8.7 Error Message "Mains Voltage < 90V"
This message is displayed when RMO100A
‟s mains voltage is below 90 V.
In this case to remove the status message you must restart RMO100A.
Figure 9-5:
Error message
"Overheat"
Figure 9-7:
Error message
"Mains Voltage < 90V"
Figure 9-4:
Error message
"Change Current"
Figure 9-6:
Error message
"Malfunction"