Product description
4
5
6
7
8
9
14
10
11
12
13
4 Filter mat
5 On/off switch
6 Power connection socket
7 Fuses
8 Operating time counter
9 Type plate
10 Compressed air supply
1)
of the ventilator
(max. 5 bar)
11 Connection
2)
for central compressed air
supply (supply pressure) in standby mode
(2.7 - 6 bar)
12 Cooling air vents
13 Unit feet / castors
14 Strain relief (for cables)
1)
Available versions: NIST connection; DISS con-
nection; quick coupling NW 7.2
2)
Available versions: NIST connection; DISS con-
nection; G1/4" internal thread
5 Operation
If the
unit is to be used as the main source of
air supply
for a ventilator, the pressure hose
(not included in the scope of delivery) must be
connected to the compressed air supply (10) of
the unit and must be connected to the ventila-
tor.
If the
unit is used in standby mode
, an
addi-
tional
pressure hose with connector (not includ-
ed in the scope of delivery) needs to be con-
nected to connector (11) of the unit and the wall
outlet of the central compressed air system. In
standby mode, the unit monitors the pressure in
the central compressed air system:
If the pressure drops below 2.7
±
0.2 bar, the
unit automatically takes over the supply of com-
pressed air to the ventilator connected to it. If
the pressure in the central compressed air sys-
tem rises above 3.2
±
0.2 bar again, the unit
switches off its supply and remains in standby
mode.
5.1 Unit details
1
2
3
1 Green control LED "Unit ready for opera-
tion"
2 Orange warning LED "Temperature too
high"
3 Cover of the unit
EN