8
1018100100L02 1907V005
Product description
4 Technical data
4.1 Basic data
Electrical data
Basic unit
Nominal pressure
bar / MPa
3 / 0.3
General data
Pressure vessel volume
L
3
Delivery quantity at 3 bar (0.3 MPa)
l/min
30
Duty cycle
%
100
Operating pressure, max.
bars
Max. 5
Switch-on/cut-off pressure(on standby)
bars
2.7
±
0.2 to 3.2
±
0.2
Supply pressure (standby mode)
bars
2.7 - 6
Safety pressure PS
bars
8
Type of protection
IP
IP20
Continuous flow
-
min. 30 L/min at 3.0 bar (the continuous
flow depends on the ambient pressure.
The actual minimum flow decreases at
the ratio of ambient pressure/1013
mbar.)
Peak flow
180 L/min for max. 0.8 sec .
Dew point reduction at operating pres-
sure
5°C at ambient temperature at
≥
30
L/min. and max. ambient temperature of
40 °C.
15°C at ambient temperature at 15
L/min. and max. ambient temperature of
40 °C
Air quality
Dust- and oil-free compressed air
Sub-micro filter
µm
≤
0.3
Noise level (at nominal pressure)
dB (A)
47
Weight
kg
40
Dimensions (L x W x H)
mm
348 x 515 x 440
Ambient conditions during ongoing
operation
temperature
°C
+5 to +40
Air pressure
hPa
500 to 1060
Rel. humidity
%
0 - 95
Classification
Medical Devices Directive (93/42/EEC)
Class IIb
EN
Summary of Contents for SICOLAB med
Page 1: ...SICOLAB med Installation and operating instructions 1018100100L02 1907V005 EN 0297...
Page 2: ......