Replace the filter housing and close.
Confirming the filter replacement
❯
Connect the mains plug.
❯
press for at least 2 seconds.
❯
press for at least 2 seconds. The unit is
now in setup mode.
Orange LED flashes
❯
Press to confirm filter replacement.
Resetting the unit to normal operation
❯
Touch.
Resetting the unit to standby mode
❯
Touch.
14 Taking out of use
14.1 Taking the unit out of use
If the compressor is not to be used for a longer
period of time, it is recommended that the unit be
properly shut down and taken out of operation.
To do so, any accumulated condensation water
must be drained from the pressure tank and from
the membrane drying unit.
The compressor must be running in order
to drain the remaining condensation
water in the water separator of the mem-
brane drying unit.
❯
Open the condensate drain valve on the mem-
brane drying unit (approx. 3 turns).
Usage
9000-610-72/30 1907V007
33
EN