5
Operation
5.1
Duo Tandem / Quattro
Tandem
1
2
3
4
7
6
5
8
9
10
11
12
13
1
Air intake filter
2
Compressor unit
3
Fine filter or sterile filter, membrane drying
unit
4
Collector tray
5
Safety valve
6
Pressure gauge/display
7
Compressed air connection (quick release
coupling)
8
Condensate drain valve
9
Pressure tank
10 Automatic/manual condensate drain valve,
membrane drying unit
11 Sinter filter for membrane drying unit
12 Intake connector
13 Controller
The compressor unit draws in atmospheric air
and compresses it without oil. It then transports
the oil-free compressed air to the membrane dry-
ing unit. The cooler and the membrane dryer
extract moisture from the compressed air. The
oil-free, hygienic and dry air is stored in the pres-
sure tank ready for use in connected devices.
All of the measurement data for the unit comes
together in the control (e.g. pressure in the pres-
sure tank, temperature of the motor windings),
where it is then evaluated. Likewise, various set-
tings (e.g. switch-on/cut-off pressure) can be
adjusted, or the unit can be connected via the
network to monitoring software.
5.2
Start-up behaviour
On compressors with an electronic controller, the
compressor units are switched on with a time
delay. The time delay depends on the operating
mode selected on the controller.
Operating mode:
– Eco: 180 s
– Balanced: 60 s
– Boost: 10 s
Alternating control:
The compressor unit with the shortest operating
time since the compressor was connected to
voltage is started. This distributes the operating
time more or less uniformly between the com-
pressor units. If the compressor is disconnected
from the voltage and is then reconnected to the
power supply, compressor unit 1 will start first
again.
Product description
9000-610-72/30 1907V007
13
EN