At maximum tank pressure, slowly open the
condensate drain valve.
Once the start-up pressure has been reached the
compressor will switch on.
On Quattro-Tandem models (2 membrane
drying units) both units must switch on.
❯
Wait until condensation water stops escaping
from the condensate drain valve on the mem-
brane drying unit.
❯
Switch off the unit - touch
for at least 2
seconds.
❯
Wait until air stops escaping from the conden-
sate drain valve (pressure tank empty).
❯
Unplug the mains plug.
❯
Close the condensate drain valve on the pres-
sure tank.
❯
Close the condensate drain valve on the mem-
brane drying unit.
❯
Disconnect the compressor from the pipe sys-
tem.
14.2
Storage of the unit
WARNING
Risk of explosion of the pressure tank
and pressure hoses
❯
The pressure tank and the pressure
hoses must be vented before they are
stored or transported.
Usage
34
9000-610-72/30 1907V007
EN