Check the firewall and release the ports, if
applicable.
Network protocols and ports
Port
Purpose
Service
45123 UDP,
45124 UDP
Unit recognition and
configuration
1900 UDP
Service detection
SSDP /
UPnP
502 TCP
Unit data
514
1)
UDP
Event protocol data
Syslog
22 TCP
Diagnosis
Telnet,
SSH
123 UDP
Time
NTP
1)
The port can vary depending on the config-
uration.
10 Adjustment options
10.1 Adjustment of the switch-
on/cut off pressure
WARNUNG
Risk of explosion of the pressure ves-
sel
The pressure vessels used in the com-
pressors are designed to withstand
continuous pressure changes of 2 bar
and can be used continuously under
these pressure changes.
❯
For load changes > 2 bar (max. per-
missible: 3 bar), comply with the maxi-
mum load change cycles specified in
the operating instructions of the pres-
sure vessel.
The pressure adjustment is performed in standby
mode.
❯
Standby button:
press for at least 2 sec-
onds.
❯
Service key:
press for at least 2 seconds.
The blue LEDs in the operating panel flash.
They are touch-sensitive and can be adjusted
accordingly.
5
6
7
8
9
bar
The blue LEDs in the operating panel flash. They
are touch-sensitive and can be adjusted accord-
ingly.
The pressure adjustment is performed in
0.5 bar increments by touching the LED.
The blue LEDs in the operating panel flash. They
are touch-sensitive and can be adjusted accord-
ingly.
❯
Touch the first flashing LED with your finger
and swipe to the required
switch on pressure
.
❯
Touch the last flashing LED with your finger and
swipe to the required
cut off pressure
.
Assembly
24
9000-610-72/30 1907V007
EN