Page
16 D-LX
100
10.5. D-LX 100 … /94 Ex and D-LX 100…/95Ex Housing for Hazardous Areas
D-LX 100 …/94 Ex
D-LX 100 …/95 Ex
LX
10
0-
08
-0
03
M20x1.5
G 1
”
1/2” NPT(F)
(Fig. 8)
Dimensional drawing: Ex-proof housing type K
Explosion protection: ..................................... D-LX 100…/94Ex:
II 2G EEx de IIC T6/T5
(PTB
approved)
D-LX 100…/95Ex:
Class I, Div. 1, Group B, C & D
(FM
approved)
Max. ambient temperature: ........................... at T5 : -20°C to +70°C (-4°F to 158°F)
at T6 : -20°C to +60°C (-4°F to 140°F)
Viewing pipe connection: .............................. G 1” pipe thread
Optical viewing angle: ................................... 6° (D-LX 100 UL: 6
°
horiz., 12
°
vert.)
Weight:........................................................... 3.8 kg (8.5 lb.)
Enclosure rating:............................................ IP 65 (rating not associated with FM logo mark)
Cable entry: ................................................... D-LX 100…/94Ex:
Cable gland M20x1.5
D-LX 100…/95Ex:
½” NPT (F) for conduit
connection
Conformity certificate:.................................... PTB 02 ATEX 1029 (D-LX 100…/94Ex ony)
Material:......................................................... AlCuMgPb
Painting:......................................................... Blue, RAL 5017 and black RAL 9005