Continued
Continued from Previous Page
Pin
Signal Name
Cable Color
Code
Description
1
Solenoid Output
BLK
Non-isolated active low digital output that is active from start of a cycle
to end of Hold (referenced to chassis ground).
2
In Cycle Status Output
WHT
Isolated output referenced to Output Common that is active from start
of a cycle to end of Hold.
3
Front Panel Lockout Input
RED
Isolated input refrenced to Input Common that locks the front panel
user interface.
4
READ RELAY NO Output
GRN
Isolated output that is referenced to READY RELAY Common that is
active when the press is ready to start a cycle.
5
READY RELAY Common
ORN
Common for READY RELAY NO Output
6
Not Used
BLU
7
Not Used
WHT/BLK
8
+22VDC Power Supply
RED/BLK
+22VDC for customer automation controls. Limited to 500mA total for
both pins.
9
+22VDC Power Supply
GRN/BLK
+22VDC for customer automation controls. Limited to 500mA total for
both pins.
10
Bad Part Status Output
ORN/BLK
Isolated output referenced to Output Common that activates when bad
part limit(s) are exceeded.
11
Top of Stroke Status Output
BLU/BLK
Isolated input referenced to Input Common that resets the Any Fault or
System Overload status outputs.
12
Not Used
BLK/WHT
13
Hold Status Output
RED/WHT
Isolated output referenced to Output Common that indicates when in
the Hold part of a weld cycle.
14
System latch Reset
GRN/WHT
Isolated input referenced to Input Common that resets the Any Fault or
System Overload status outputs.
15
Input Common
BLU/WHT
Isolated input Common
16
+22VDC Power Ground
BLK/RED
+22VDC Power Ground (also chassis ground)
17
+22VDC Power Ground
WHT/RED
+22VDC Power Ground (also chassis ground)
18
+22VDC Power Ground
ORN/RED
+22VDC Power Ground (also chassis ground)
19
Ultrasound Active Status
BLU/RED
Isolated output referenced to Output Common that indicates ultrasound
is being output.
20
System Ready Status
RED/GRN
Isolated output referenced to Output Common that is active when the
press is ready to start a cycle.
21
Any Fault Status
ORN/GRN
Isolated output referenced to Output Common that is active when the
any fault/alarm is detected.
22
Output Common
BLK/WHT/RED Isolated Output Common.
23
Output Common
WHT/BLK/RED Isolated Output Common.
24
U/S Activate Common
RED/BLK/WHT Isolated U/S Activate/Cycle Start.
25
U/S Activate/Cycle Start
Input
GRN/BLK/WHT Isolated input referenced to Isolated U/S Activate/Cycle Start Input
Common that starts a cycle when activated.
Table 3-I
System Input/Output Connector Signals (J1)
Page 21
Dukane Manual Part No. 403-594-02
Section 3 - Installation