oFFXX
Auto Power Off Time
Defines the number of minutes (from 1 to 30) before the automa c switch-off in
case of constant pressure. The power off me starts working if the instrument does
not detect pressure varia ons greater than 10% of full scale value.
Modify the parameter using the
-keys.
↓↑
contX
Con nuous data transmission mode via USB: on/off
X = 0: the data transmission is obtained on request
X = 1: the pressure value is transmi ed con nuously every 100 ms.
Modify the parameter using the
-keys.
↓↑
Lt XX
Backlight ac va on me
This parameter defines the switching me of the backlight ac vated when the
SET key is pressed for a short me. The me can be set between 1 s and 99 s.
By se ng 0, the backlight is turned off. Ac va ng the backlight will result in
higher ba ery consump on, so it is be er to deac vate this func on when it is
not used.
ZEroX
Enable the ZERO func on (only for
pressure reading instruments)
gauge
X = 0: the zero func on is disabled (
for
ranges)
recommended
absolute
X = 1: enables the zero func on.
6.5 Data Logger paramters: only for instruments with op onal data logger func onality
AUtdL
enables AUTOMATIC datalogger capture mode (based on t1 and t2 parameters).
MAndL
enables MANUAL capture mode, controlled by the operator.
t On
Enables recording of both, pressure and temperature values
t OFF
Enables recording of just pressure values
t1
Defines the me between two consecu ve capture intervals in automa c mode.
You can set: hh = hours / mm = minutes / ss = seconds
hh.mm.ss
Minimum value 1 sec., maximum value 24 hours.
Example:
1.30.05
=
1 hour, 30 minutes, 5 seconds
t2
Defines the global me of recording cycle dura on.
You can set: gg = days / hh = hours / mm = minutes
ggg.hh.mm
Minimum value 1 minute, maximum value 365 days (limited by instrument
memory capacity).
Example:
00.24.30
=
24 hours, 30 minutes