DRUCK TEMPERATUR
&
Leitenberger GmbH •
Bahnhofstr. 33 • D-72138 Kirchentellinsfurt • GERMANY
Tel. +49 (0) 7121-90920-0 • Fax +49 (0) 7121-90920-99 • [email protected] •
DT
www.druck-temperatur.de
MANUAL
TLDMM
Referezdruckmessgerät
2.0
TLDMM
reference pressure gauge
2.0
39
6.6 Se ng of Date and Time
Press the keys SET and PEAK simultaneously for a few seconds.
P0000
Set the
with the arrow-keys and confirm with SET.
password 8321
d1
Allows to set the YEAR. Modify the parameter using the arrow-keys and confirm
with SET key to go to the next parameter.
d2
Allows to set the MONTH. Modify the parameter using the arrow-keys and confirm
with SET key to go to the next parameter.
d3
Allows to set the DAY. Modify the parameter using the arrow-keys and confirm
with SET key to go to the next parameter.
d4
Allows to set the HOUR. Modify the parameter using the arrow-keys and confirm
with SET key to go to the next parameter.
d5
Allows to set the MINUTE. Modify the parameter using the arrow-keys and confirm
with SET key to finish se ng the date and me.
If no parameter is changed, the internal date/ me on the instrument is not modified. During the
date/ me update, the seconds are automa cally set to zero.
6.7 Temperature display unit
Press simultaneously and hold for a few seconds the keys SET and PEAK.
P0000
Select the password
using the Arrow-Down- and Arrow-Up-Keys
0033
and confirm by pressing SET-key.
Unit
Change the temperature unit from °C to °F or reverse using the Arrow-Down-
or Arrow-Up-Key. Confirm with SET key.
6.8 Key lock func on
Press simultaneously and hold for a few seconds the keys SET and PEAK.
P000
Select the password
using the Arrow-Down- and Arrow-Up-Keys
0301
and confirm by pressing SET-key.
LOC X
X = 0: Key block func on disabled; X = 1: Key block func on enabled,
shown on display. Keys SET, ZERO and PEAK are disabled. This func on allows to
avoid that not authorized personnel can modify some parameters & func ons.