CAUTION!
The temperature range over which the Lithium-Ion ba ery can be discharged is
0...+50°C. Use of the Lithium-Ion ba ery outside of this temperature range may
affect the performance of the ba ery or may reduce its service life.
6. Opera on
6.1 Parameters programming
The instrument can be programmed through the parameters menu:
• Pressure unit
• Digital filter
• Display resolu on
• Auto-Power-Off me
• Indica on of measured medium temperature (toggle mode with indica on of pressure)
• If your instrument is equipped with op onal data logger func onality:
- Storage interval
- Max. storage dura on
- Data selec on (pressure only, or pressure plus medium temperature)
• Dura on of display backlight
• Baud rate USB
6.2 Keys descrip on
Key with 4 func ons:
• pushed: Switch on the instrument
• pushed for 3 seconds: Enter the parameters menu
• pushed for 5 seconds: Switch off the instrument
• If enabled, pushing this key switches on the display backlight for
programmed period of me
Key with 4 func ons:
• pushed during measurement, for 3 seconds: performs the ZERO of the display
up to 50% of full scale value. The ZERO func on does not affect the bargraph
display.
• pushed during measurment, for 6 seconds: switches off the ZERO func on.
• pushed in PEAK mode: rese s the stored max. value.
• pushed in parameters menu: decreases indicated value.
Key with 4 func ons:
• pushed during measurement, for 2 seconds: ac vates Max. PEAK mode.
• pushed during measurement, for 4 seconds: ac vates Min. PEAK mode.
• pushed during measurement, for 6 seconds: medium temperature indica on.
if again pushed for 6 seconds: back to pressure value indica on.
• pushed in parameters menu: increases indicated value.