![DRUCK & TEMPERATUR Leitenberger LR-Cal LRT 1000 Operating Manual Download Page 36](http://html1.mh-extra.com/html/druck-and-temperatur-leitenberger/lr-cal-lrt-1000/lr-cal-lrt-1000_operating-manual_2533109036.webp)
DRUCK & TEMPERATUR Leitenberger GmbH •
Bahnhofstr. 33 • D-72138 Kirchentellinsfurt • GERMANY
Tel. +49 (0) 7121-90920-0 • Fax +49 (0) 7121-90920-99 • [email protected] •
• www.druck-temperatur.de
www.LR-Cal.net
MANUAL
XP100 und
LRT 1000
LR-
Cal
XP100 and
LRT 1000
LR-
Cal
36
Ba ery change
Change the ba eries when the message „Ba lo“ is displayed upton switch-on or the device can
no longer be switched on. See „Inserng the ba eries“ on page 26.
Available Pt 100 sensors
An overview of the currently available Pt 100 sensors is provided below.
At model
a precision ceramic sensor is already included in the standard supply.
LRT 1000
LR-
Cal
For model
following sensors are available (connecon cable 2 m length or 10 m length
XP100
must be ordered seperately):
Order-Code
3120.520
Immersion probe, short, 150 x 3 mm, -40...+400°C
Order-Code
3120.530
Immersion probe, long, 300 x 3 mm, -40...+400°C
Order-Code
3120.540
Immersion/penetraon probe, long, 300 x 4 mm
Order-Code
3120.550
Stainless steel food penetraon probe
Order-Code
3120.600
Surface probe, -50...+400°C
NOTE:
DRUCK & TEMPERATUR Leitenberger GmbH improves
and broadens the range of available sensors. Informaon
on addional, new sensors can be found on the website
www.druck-temperatur.de
Connecng third-party-sensors
When connecng third-party sensors to the device
(model XP100), only use the original plug-in connector
(order-code 3120.50). The maximally permissable connecon cross-secon is 0.75 mm².
Please observe the following connecon diagram:
Pin 1 = Force High (+), Pin 2 = Sense High (+),
Pin 3 = Sense Low (-), Pin 4 = Force Low (-)
NOTE: Sensors must be connected using the 4-wire technique,
the 3-wire technique is not supported.
3120.
540
3120.
520
3120.
530
3120.
600