b)
If W >= 1
W = aH[W - 1] + bH[W - 1]² + cH[W - 1]³ + dH[W - W Al)]²
r
Default se ng:
aH = 0; cH = 0; dH = 0; W-Al = 3.3760086
„Measuring mode“ screen
fig. A
fig. B
(40)
Indicates the current temperature value.
(41)
Displays informaon on the current measuring mode and
the characterisc curve.
(42)
This symbol indicates, that the displayed measured values
under (43) are calculated (e.g. minimum/maximum
measured value).
(43)
Indicates the calculated measured values according to the
selecon (e.g. minimum/maximum measured value).
(44)
Sets the measured values displayed under (43) and the
duraon back to zero.
(45)
Specifies how the measured values under (43) are shown
(see fig. ):
B
Minimum:
Always shows the smallest detected
measured value from a measuring period.
Maximum:
Always shows the largest detected
measured value from a measuring period.
Average:
Shows the average value of all measured
values which have been detected so far
from a measuring period.
Hold:
Pauses the current detected measured
value and shows it connually (when this
opon is selected).
Off:
Switches the display (42) off.
(46)
Carries out an automac measurement for the duraon of
an already specified recording interval. The recording
interval can be specified in the following screen (see fig.
C
and fig.
on the next page).
D
(47)
Saves the currently displayed measured values as single
measurement in the archieve with date and me stamp.
40
42
43
44
45
46
47