9.2 Points Calibraon procedure (kg / t / daN / kN)
• Move the measuring unit in the desired scale.
• Bring the load of the hydraulic machine to 0 (zero).
• Press together SET and PEAK keys, to enter into the password selecon menu.
P0000
Set the password
using the Arrow-Up/Arrow-Down keys, then confirm with
3464
the SET key.
P0 c
The label
appears, then the offset value. Press the ZERO key to reset, then
P0 c
press the SET key to confirm.
P1 c
The label
appears, press the SET key. Bring the hydraulic machine to the
P1 c
20% of the load and change the value on the display with Arrow-Up/Arrow-Down
keys. Press SET key to confirm.
P2 c
The label
appears, press the SET key. Bring the hydraulic machine to the
P2 c
40% of the load and change the value on the display with Arrow-Up/Arrow-Down
keys. Press SET key to confirm.
P3 c
The label
appears, press the SET key. Bring the hydraulic machine to the
P3 c
60% of the load and change the value on the display with Arrow-Up/Arrow-Down
keys. Press SET key to confirm.
P4 c
The label
appears, press the SET key. Bring the hydraulic machine to the
P4 c
80% of the load and change the value on the display with Arrow-Up/Arrow-Down
keys. Press SET key to confirm.
P5 c
The label
appears, press the SET key. Bring the hydraulic machine to the
P5 c
100% of the load and change the value on the display with Arrow-Up/Arrow-Down
keys. Press SET key to confirm.
dP
The label
appears, via Arrow-Up/Arrow-Down keys select the decimal point.
dP
Confirm by pressing the SET key.
End
The label
appears. The adjustment/calibraon is finished and the instrument
End