![DRUCK & TEMPERATUR Leitenberger LR-Cal DFP Operating Manual Download Page 24](http://html1.mh-extra.com/html/druck-and-temperatur-leitenberger/lr-cal-dfp/lr-cal-dfp_operating-manual_2533104024.webp)
DRUCK & TEMPERATUR Leitenberger GmbH •
Bahnhofstr. 33 • D-72138 Kirchentellinsfurt • GERMANY
Tel. +49 (0) 7121-90920-0 • Fax +49 (0) 7121-90920-99 • [email protected] •
• www.druck-temperatur.de
www.LR-Cal.net
MANUAL
TLDMM + TLDMM-A01
TLDMM-A02 + DFP
24
9. Force Unit adjustment - only model
DFP (kg / t / daN / kN)
LR-
Cal
The instruments model
are supplied as a normal
DFP
LR-
Cal
digital pressure gauge, calibrated in unit „bar“.
It is up to the installer to set the full scale in the
force unit used, by using the following formular:
F[N] = P x D x
/ 40
²
P
P
= pressure in bar
D
= cylinder diameter in mm
Conversion from force to weight with following formular:
F[kg] = F[N] / 9,80665
Measurement conversion in the other available units is
automac.
9.1 Force unit adjustment procedure (kg / t / daN / kN)
• Select the desired scale/unit on the instrument.
• Bring the load of the hydraulic machine to 0 (zero).
• Press together SET and PEAK keys to enter into passwort selecon menu.
P0000
Set the password
using the Arrow-Up and Arrow-Down keys and confirm
4254
with SET key.
ForFS
string appears and aer 2 seconds the Full Scale expressed in
is shown.
ForeFS
bar
Set the Full Scale in the selected unit via Arrow-Up / Arrow-Down keys and press
SET key to confirm.
dP
string appears. In this phase the decimal point has to be set. Confirm using the
dp
SET key, move the decimal point using the Arrow-Up/Arrow-Down keys and
confirm using the SET key.
End
Ende der Prozedur.
Adjustment cancellaon:
To delete the performed adjustment, enter password
and confirm with SET key.
4256
When
measurement unit is selected, instrument works with the original Full Scale.
bar