5

Symbol

Item

Value

Description

Factory

set value

AL

Setting

value for

lower

deviation

alarm

-50.

100.0

0 means no alarm, 1 menas lower

deviation

alarm.

When

the

temperature

is

less

than

(temperature set value-AL)

, the

refrigeration will be turned off. The

abnormal indicator lights up and the

buzzer sounds.

6.0

AH

Setting

value for

upper

deviation

alarm

-50.

100.0

Upper deviation alarm, when the

temperature exceeds (temperature

set value + AL)

, turn off the

heating. When the temperature is

lower than (temperature setting

value-AL)

,

turn

off

the

refrigeration. The abnormal indicator

light is on and the buzzer sounds.

6.0

AP

Max.

Protection

temp

Max.measur

ement temp.

When the temp. exceeds AP, the

heating will be cut off. When the

temp. exceeds the AP parameter for

2 minutes, the buzzer keeps beeping

and all outputs are turned off. It must

be powered off to recover.

120.0

P

Heating

proportional

band

1.0

60.0

Proportional adjustment, the smaller

the Pu, the faster the response

speed.

6.5

I

Heating

integration

time

1

3600S

Integral action time constant, the

smaller the Iu, the stronger the ability

to correct the static error.

240

d

Heating

differential

time

0

3600S

Differential action time constant, the

larger the du, the stronger the ability

to prevent over-temperature.

180

Ar

Heating

overshoot

suppression 0

100

Used to suppress temperature

overshoot.

80

t

Heating

control cycle 1

99S

Control heating output cycle

5S

ct

Compressor

protection

delay

(0

9999)S

Compressor delay protection time,

the two start-up time is ≥CT seconds;

Note that when Ct=0, there is no

refrigeration function.

120S

CL

Turn

off

refrigeration

(0

10.0)

In

the

switching

mode,

the

measured temp. <(set tempe

CL), turns off the refrigeration. Used

to initialize parameters. Note: when

CL=0, manual mode is valid.

0

CH

Turn

on

efrigeration

(0

10.0)

In the switching mode, the measured

temp.> (set temp. + CH) ,turns on the

refrigeration.

Used

to

initialize

parameters. Note: Manual mode is

valid when CH=0

0

pb

Zero

adjustment

of

temperature

(-199

999)

Correction around zero degrees,

can be used to correct the error

generated during the temp. sensor

measurement.

PB=Mercury

thermometer reading value-current

measurement value.

0

Summary of Contents for DW-LBI Series

Page 1: ...gy Limited Chongqing Drawell Instrument Co Ltd Shanghai Drawell Scientific Instrument Co Ltd Chongqing Center Suite 2705 Building No 12 Shiyou Road No 1 Yuzhong District Chongqing China Shanghai Offic...

Page 2: ...I Application 1 III Technical Specifications 2 IV Installation requirements 2 V Operation Instruction 3 VI Wiring diagram 6 VII Malfunction handling 7 VIII Transportation and storage 7 IX After sales...

Page 3: ...he inside chamber can be cleaned regularly with dry cloth When the set temperature is lower than the room temperature the over temperature alarm will be generated if the door opening time is too long...

Page 4: ...be grounded The equipment shell should be grounded reliably and placed in a sheltered place from the sun cool and ventilated V Operation instruction Model DW LBI 80 DW LBI 150 DW LBI 200 DW LBI 250 D...

Page 5: ...always on for illumination flashing for UV disinfection 16 Heating indicator light on when heated 17 The refrigeration indicator lights when there is refrigeration output and flashes when refrigeratio...

Page 6: ...1 0 minutes 1 hour 0 Fn High and low fan switch 0 100 Start the low speed fan when the temperature is between the set value Fn and the set value Fn start the high speed fan when the temperature is out...

Page 7: ...ng integration time 1 3600S Integral action time constant the smaller the Iu the stronger the ability to correct the static error 240 d Heating differential time 0 3600S Differential action time const...

Page 8: ...value current measurement value current measurement value 0 cb Ambient temperature correction 60 100 0 cb actual ambient temp currently measured ambient temp 0 rl Temp lower limit setting 60 100 0 Th...

Page 9: ...e fan The meter is not corrected Correct the meter Note Maintenance operations shall be carried out by qualified personnel Please turn off the power supply before repairing VIII Transportation and sto...

Page 10: ...o change the data in the manual without prior notice The company has the final interpretation right 1 Biochemical incubator Machine 1 Set 2 User Manual Document 1 Copy 3 Fuse Spare parts 1pc 4 This pa...

Page 11: ...CO Ltd Add Suite 2705 Building No 12 Shiyou Road No 1 Yuzhong District Chongqing China Tel 0086 023 63268643 Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1117 Lane561 XiuChuan Rd PuDong New...

Reviews: