5
Symbol
Item
Value
Description
Factory
set value
AL
Setting
value for
lower
deviation
alarm
-50.
~
100.0
℃
0 means no alarm, 1 menas lower
deviation
alarm.
When
the
temperature
is
less
than
(temperature set value-AL)
℃
, the
refrigeration will be turned off. The
abnormal indicator lights up and the
buzzer sounds.
6.0
℃
AH
Setting
value for
upper
deviation
alarm
-50.
~
100.0
℃
Upper deviation alarm, when the
temperature exceeds (temperature
set value + AL)
℃
, turn off the
heating. When the temperature is
lower than (temperature setting
value-AL)
℃
,
turn
off
the
refrigeration. The abnormal indicator
light is on and the buzzer sounds.
6.0
℃
AP
Max.
Protection
temp
Max.measur
ement temp.
When the temp. exceeds AP, the
heating will be cut off. When the
temp. exceeds the AP parameter for
2 minutes, the buzzer keeps beeping
and all outputs are turned off. It must
be powered off to recover.
120.0
℃
P
Heating
proportional
band
1.0
~
60.0
℃
Proportional adjustment, the smaller
the Pu, the faster the response
speed.
6.5
℃
I
Heating
integration
time
1
~
3600S
Integral action time constant, the
smaller the Iu, the stronger the ability
to correct the static error.
240
d
Heating
differential
time
0
~
3600S
Differential action time constant, the
larger the du, the stronger the ability
to prevent over-temperature.
180
Ar
Heating
overshoot
suppression 0
~
100
%
Used to suppress temperature
overshoot.
80
%
t
Heating
control cycle 1
~
99S
Control heating output cycle
5S
ct
Compressor
protection
delay
(0
~
9999)S
Compressor delay protection time,
the two start-up time is ≥CT seconds;
Note that when Ct=0, there is no
refrigeration function.
120S
CL
Turn
off
refrigeration
(0
~
10.0)
℃
In
the
switching
mode,
the
measured temp. <(set tempe
CL), turns off the refrigeration. Used
to initialize parameters. Note: when
CL=0, manual mode is valid.
0
CH
Turn
on
efrigeration
(0
~
10.0)
℃
In the switching mode, the measured
temp.> (set temp. + CH) ,turns on the
refrigeration.
Used
to
initialize
parameters. Note: Manual mode is
valid when CH=0
0
pb
Zero
adjustment
of
temperature
(-199
~
999)
Correction around zero degrees,
can be used to correct the error
generated during the temp. sensor
measurement.
PB=Mercury
thermometer reading value-current
measurement value.
0