background image

2

Note

:Performance parameters are tested under non-load conditions: ambient

temperature is 20°C, relative humidity is 50% RH.

IV. Installation requirements

1

) The instrument should be placed in a constant temperature environment, and the

recommended temperature difference is 23 ± 5

;

2

) The distance between the equipment and the wall must be more than 10cm and the

button operation position shall be reserved on the right side.

3) Avoid direct sunlight and keep away from heating equipment;

4) Place the equipment on the horizontal ground, lock the front caster to make the box flat.

5) Turn on the 220V/50Hz power supply and it should be grounded. The equipment

shell should be grounded reliably and placed in a sheltered place from the sun, cool

and ventilated.

V. Operation instruction

Model

DW-LBI-

80

DW-LBI-150

DW-LBI-200

DW-LBI-250

DW-LBI-300

DW-LBI-400

Chamber Volume

80L

150L

200L

250L

300L

400L

Temperature Range

0~60

Display Resolution

0.1

Temperature Stability

±1

Temperature
Uniformity

±1

Timing Range

0~9999min

Power Rating

W

180

250

300

350

400

550

Refrigerant

R134a

Power Supply

220V , 50Hz

Continuous Operation

Long continuous operation

Chamber
Size(W×D×H)cm

40×37×55

45×42×85

45×45×100

48×49×107

52×50×117

58×54×127

Exterior
Size(W×D×H)cm

54×57×107

59×62×137

59×64×152

62×68×159

66×69×169

72×74×179

Package Size
(W×D×H)cm

70×73×126

75×78×156

75×80×171

78×84×178

82×85×188

88×90×198

Net/Gross Weight

kg

50/85

90/120

95/138

103/147

115/157

135/170

Summary of Contents for DW-LBI Series

Page 1: ...gy Limited Chongqing Drawell Instrument Co Ltd Shanghai Drawell Scientific Instrument Co Ltd Chongqing Center Suite 2705 Building No 12 Shiyou Road No 1 Yuzhong District Chongqing China Shanghai Offic...

Page 2: ...I Application 1 III Technical Specifications 2 IV Installation requirements 2 V Operation Instruction 3 VI Wiring diagram 6 VII Malfunction handling 7 VIII Transportation and storage 7 IX After sales...

Page 3: ...he inside chamber can be cleaned regularly with dry cloth When the set temperature is lower than the room temperature the over temperature alarm will be generated if the door opening time is too long...

Page 4: ...be grounded The equipment shell should be grounded reliably and placed in a sheltered place from the sun cool and ventilated V Operation instruction Model DW LBI 80 DW LBI 150 DW LBI 200 DW LBI 250 D...

Page 5: ...always on for illumination flashing for UV disinfection 16 Heating indicator light on when heated 17 The refrigeration indicator lights when there is refrigeration output and flashes when refrigeratio...

Page 6: ...1 0 minutes 1 hour 0 Fn High and low fan switch 0 100 Start the low speed fan when the temperature is between the set value Fn and the set value Fn start the high speed fan when the temperature is out...

Page 7: ...ng integration time 1 3600S Integral action time constant the smaller the Iu the stronger the ability to correct the static error 240 d Heating differential time 0 3600S Differential action time const...

Page 8: ...value current measurement value current measurement value 0 cb Ambient temperature correction 60 100 0 cb actual ambient temp currently measured ambient temp 0 rl Temp lower limit setting 60 100 0 Th...

Page 9: ...e fan The meter is not corrected Correct the meter Note Maintenance operations shall be carried out by qualified personnel Please turn off the power supply before repairing VIII Transportation and sto...

Page 10: ...o change the data in the manual without prior notice The company has the final interpretation right 1 Biochemical incubator Machine 1 Set 2 User Manual Document 1 Copy 3 Fuse Spare parts 1pc 4 This pa...

Page 11: ...CO Ltd Add Suite 2705 Building No 12 Shiyou Road No 1 Yuzhong District Chongqing China Tel 0086 023 63268643 Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1117 Lane561 XiuChuan Rd PuDong New...

Reviews: