- 20 -
5) Wavelength Calibration
After some time, the energy of lights would drop off and there
will be some but slight influence on the test result. In this case,
users can reset the wavelength to compensate. We suggest the
users to reset the wavelength every one or two months.
Remove all the blocks in the light path and close the lid of the compartment,
press Arrow Key to choose “Calibrate WL.” and press
ENTER
√Setup
Date & Time
Dark Current
Calibrate WL.
Calibrating WL.
to begin the
calibration
(
Fig. 4-28
)
. The lid of the compartment should not be opened
during the calibration. It returns automatically after calibration.
ENTER
√Setup
Date & Time
Dark Current
Calibrate WL.
Calibrating WL.
Fig. 4-28
6) Lamps Change
(Input the lamps changing wavelength point)
The instrument permits the users to set the lamps changing
wavelength point. In the range of 300-400nm, users can
choose the lamp freely.
Press Arrow Key to choose “Lamp Change” and press
√Setup
Dark Current
Calibrate WL.
Lamps Change
ESC
ENTER
Lamps Change
WL=500.0nm
Please Input WL.
WL=
to go into
corresponding setting interface (Fig.4-29). Input the value of lamps changing
wavelength point by the Arrow keypad
(
300—400nm
)
,
then press
√Setup
Dark Current
Calibrate WL.
Lamps Change
ESC
ENTER
Lamps Change
WL=500.0nm
Please Input WL.
WL=
to
confirm and return.
√Setup
Dark Current
Calibrate WL.
Lamps Change
ESC
ENTER
Lamps Change
WL=500.0nm
Please Input WL.
WL=
Fig. 4-29