- 19 -
ESC
ENTER
√Setup
D2 Lamp On
W Lamp On
Date & Time
√W Lamp
On
√
Off
Fig. 4-25
3) Set Date & Time
Press Arrow Key to choose “Date & Time” and press
ESC
ENTER
√Setup
D2 Lamp On
W Lamp On
Date & Time
Date & Time
Time: 12: 00: 00
Date: 2008/08/08
Week Friday
to go into
setting interface.
(
Fig. 4-26
)
Use Arrow Key to choose setting item and
input the value by Arrow keypad.
(
Hour
:
0-23
,
Minute
:
0-59
,
Second
:
0-59
,
Year
:
0000-9999
,
Month
:
1-12
,
Day
:
1-31, Week: 1-7
),
press
ESC
ENTER
√Setup
D2 Lamp On
W Lamp On
Date & Time
Date & Time
Time: 12: 00: 00
Date: 2008/08/08
Week Friday
to
return.
ESC
ENTER
√Setup
D2 Lamp On
W Lamp On
Date & Time
Date & Time
Time: 12: 00: 00
Date: 2008/08/08
Week Friday
Fig. 4-26
4) Calibrate Dark Current
When the working condition changes, A dark current
calibration is required before any measurement.
Use Arrow key to choose “Dark Current” and press
ENTER
√Setup
W Lamp On
Date & Time
Dark Current
Getting Dark...
to confirm. (Fig. 4-27
)
Note: Remove the cuvettes from the light path before this action, don’t
open the lid of the compartment during the course of calibration.
Press
ENTER
√Setup
W Lamp On
Date & Time
Dark Current
Getting Dark...
to return when it finishes.
ENTER
√Setup
W Lamp On
Date & Time
Dark Current
Getting Dark...
Fig. 4-27