- 16 -
II Coefficient Method
:
(1) Go into
“
Coefficient
”
method.
Press Arrow Key to choose “Coefficient”, press
√Quantitative
Standard Curve
Coefficient
Unit
Coefficient
0.000 Abs
0.000 mg/ml
500.0nm
ESC
ENTER
Coefficient
ESC
SET
Coefficient K
Coefficient B
to go into
pre-test interface, then press
√Quantitative
Standard Curve
Coefficient
Unit
Coefficient
0.000 Abs
0.000 mg/ml
500.0nm
ESC
ENTER
Coefficient
ESC
SET
Coefficient K
Coefficient B
to go into the parameter setting
interface.
(
Fig. 4-19
)
;
√Quantitative
Standard Curve
Coefficient
Unit
Coefficient
0.000 Abs
0.000 mg/ml
500.0nm
ESC
ENTER
Coefficient
ESC
SET
Coefficient K
Coefficient B
Fig. 4-19
(2) Parameter Setting
★
Use Arrow Key to move the cursor on “Coefficient K” and press
√Coefficient
Coefficient K
Coefficient B
Coefficient K
C=K x A+B
K=0.000
K=
ESC
ENTER
to go into K setting interface.
(
Fig. 4-20
)
;
Input K and press
√Coefficient
Coefficient K
Coefficient B
Coefficient K
C=K x A+B
K=0.000
K=
ESC
ENTER
to confirm
,
the system will return the former interface
automatically.
√Coefficient
Coefficient K
Coefficient B
Coefficient K
C=K x A+B
K=0.000
K=
ESC
ENTER
Fig. 4-20
★
Input the value of B in the same way, press
Coefficient
0.000 Abs
0.000 mg/ml
500.0 nm
Quantitative
WL=500.0nm
Please Input WL.
WL=
ESC
GOTOλ
to go back to
the pre-test interface.
(
Fig. 4-19
)
(3) Set Wavelength
Press
Coefficient
0.000 Abs
0.000 mg/ml
500.0 nm
Quantitative
WL=500.0nm
Please Input WL.
WL=
ESC
GOTOλ
to go into wavelength setting interface
,
input the
wavelength value by the Arrow keypad and press
Coefficient
0.000 Abs
0.000 mg/ml
500.0 nm
Quantitative
WL=500.0nm
Please Input WL.
WL=
ESC
GOTOλ
to
confirm. (Fig. 4-21
)