- 15 -
★
Move the cursor on the equation that you need and press
ESC
ENTER
No. Curve Data
1 C=1.852A+2.010
2 C=1.000A+0.000
0.000 0.800
28.00
0.000
r=1.000
C=1.852A+2.010
to
confirm. Then the corresponding curve will display on the screen.
(Fig. 4-16)
ESC
ENTER
No. Curve Data
1 C=1.852A+2.010
2 C=1.000A+0.000
0.000 0.800
28.00
0.000
r=1.000
C=1.852A+2.010
Fig. 4-16
★
Press
Standard Curve
ESC
ENTER
Create Curve
Load Curve
Delete Curve
No. Curve Data
1 C=1.852A+2.010
2 C=1.000A+0.000
to go into sample continuous measure interface and
measure samples. (Fig. 4-14)
Note: When you reload the curves, the wavelength will be set at the
point where you creat the Standard Curve automatically.
(7) Delete Curve
★
Move the cursor on “Delete Curve” and press
Standard Curve
ESC
ENTER
Create Curve
Load Curve
Delete Curve
No. Curve Data
1 C=1.852A+2.010
2 C=1.000A+0.000
to confirm.
Then the system goes into the curve equation choosing interface. (Fig.
4-17)
Standard Curve
ESC
ENTER
Create Curve
Load Curve
Delete Curve
No. Curve Data
1 C=1.852A+2.010
2 C=1.000A+0.000
Fig. 4-17
★
Move the cursor on the curve equation that you don’t need again
and press
ESC
ENTER
No. Curve Data
1 C=1.852A+2.010
2 C=1.000A+0.000
Delete Curve
YES
NO
, then system will ask you to confirm again.(Fig.
4-18)
ESC
ENTER
No. Curve Data
1 C=1.852A+2.010
2 C=1.000A+0.000
Delete Curve
YES
NO
Fig. 4-18
★
Move the cursor on “Yes” and press
√Quantitative
Standard Curve
Coefficient
Unit
Coefficient
0.000 Abs
0.000 mg/ml
500.0nm
ESC
ENTER
Coefficient
ESC
SET
Coefficient K
Coefficient B
to confirm, then the
curve equation will be deleted. If you don’t want to delete it, just
choose “NO” or press
√Quantitative
Standard Curve
Coefficient
Unit
Coefficient
0.000 Abs
0.000 mg/ml
500.0nm
ESC
ENTER
Coefficient
ESC
SET
Coefficient K
Coefficient B
to return.