- 12 -
ESC
ENTER
√Quantitative
Standard Curve
Coefficient
Unit
1. %
2. ug/l
3. mg/l
4. g/l
5. ml/l
6. mg/ml
7. ug/ml
8. None
Fig. 4-7
Step 3: Choose Methods
There are two methods to be selected: Standard Curve and Coefficient.
I. Standard Curve
:
(1) Go into
“
Standard Curve
”
Interface
In “Quantitative” menu, press Arrow Key to choose “Standard Curve”;
Press
ENTER
to go into sub menu.
(
Fig. 4-8
);
Here you can create
a new curve, or you can load a saved curve. If you want to delete a
saved curve, just choose
“
Delete Curve
”
.
Standard Curve
ESC
ENTER
√Quantitative
Standard Curve
Coefficient
Unit
Create Curve
Load Curve
Delete Curve
Fig. 4-8
(2) Set Wavelength
★
Move the cursor on
“
Create Curve
”
, the press
ENTER
to go into
the sample quantity setting interface. (Fig. 4-9)
Standard Curve
ESC
ENTER
Create Curve
Load Curve
Delete Curve
500.0nm 0.000Abs
Input Stds. No.
No.= 3
Fig. 4-9
★
Press
Quantitative
WL=500.0nm
Please Input WL.
WL=
GOTOλ
500.0nm 0.000Abs
Input Stds No:
No.= 3
to go into the wavelength setting interface
,
input the
WL. Value by the Arrow Keypad and press
Quantitative
WL=500.0nm
Please Input WL.
WL=
GOTOλ
500.0nm 0.000Abs
Input Stds No:
No.= 3
to confirm.
(
Fig.
4-10
)
;