Dräger X-am 2500
27
Operation
1. Switching off the instrument: Press and hold OK key and [+] key
simultaneously.
2. Loosen the screw on the power pack and remove the power pack.
Battery holder (order no. 83 22 237): Replace alkaline batteries or
NiMHy rechargeable batteries. Ensure correct polarity.
NiMH power pack T4 (Type HBT 0000) / T4 HC (Type HBT 0100):
replace complete power pack.
3. Insert the power pack into the instrument and tighten the screw,
the instrument switches on automatically.
4.9.2
Charge instrument with NiMH power pack T4
(Type HBT 0000) / T4 HC (Type HBT 0100)
Insert the switched off instrument into the charger module.
Display LED on the charger module:
To protect the battery charge only in the temperature range of 5 to
35 °C. Outside this temperature range, the charging process is
automatically interrupted and automatically continued after the
temperature range has been reached again. The charging time is
typically 4 hours. A new NiMH power pack reaches its full capacity
after three complete charge / discharge cycles. Never store the
instrument for extended periods without being connected to a power
source (maximum of 2 months) because the internal buffer battery will
drain.
4.9.3
Carry out manual bump test
1. Prepare a test gas cylinder, the volume flow must be 0.5 l/min and
the gas concentration must be higher than the alarm threshold
concentration that is to be tested.
2. Connect the test gas cylinder with the calibration cradle (order no.
83 18 752).
3. Switch on the instrument and insert it into the calibration cradle –
press downwards until it engages.
WARNING
Risk of explosion!
The gas detector must only be operated with the ABT 0100
battery holder.
WARNING
Explosion hazard! To reduce the risk of ignition of a flammable
or explosive atmosphere, strictly adhere to the following
warning statements:
Do not charge underground or in explosion hazard areas!
The chargers are not designed in accordance with the
regulations for fire damp and explosion protection.
Charge NiMH power packs T4 (type HBT 0000) or T4 HC
(type HBT 0100) with the appropriate Dräger charger.
Ambient temperature during charging: 0 to +40 °C.
Charge
Malfunction
Full
!
!
NOTICE
The automatic bump test with the Bump Test Station is
described in the Technical Handbook.
WARNING
CSA requirement: carry out a bump test before use. It should
be carried out in the measuring range 25-50 % of the full
scale value, whereby the displayed measured value may
deviate from the actual measured value by 0-20 %. Accuracy
may be corrected via calibration.
CAUTION
Never inhale the test gas. Health hazard!
Observe the hazard warnings of the relevant Safety Data
Sheets.
i
i
!
!
Summary of Contents for X-AM 2500
Page 2: ......
Page 230: ...230 Dr ger X am 2500 1 245 5 Dr ger Dr ger Dr ger Dr ger 1 1 www draeger com ifu i i i i...
Page 237: ...Dr ger X am 2500 237 CatEx 125 PR Dr gerSensor CatEx 125 PR Gas...
Page 238: ...238 Dr ger X am 2500 4 6 4 6 1 A1 A1 O2 A1 A1 A1 A2 OK 4 6 2 A2 A2 O2 A1 A2 OK CatEx CatEx i i...
Page 247: ......