Configuration
2.3
Approvals
Copies of the name plate and the declaration of conformity are
provided in the enclosed supplementary documentation (order no.
90 33 890).
Do not stick anything on the name plate on the gas detector. The
technical suitability tests are valid for the X-am 2500 gas detection
instrument and the calibration cradle. The explosion-protection
approvals are only valid for the X-am 2500 gas detection instrument;
the calibration cradle must not be used in the Ex zone.
The BVS 10 ATEX E 080 X technical suitability test is based on the
adjustment with the target gas.
3
Configuration
To individually configure an instrument with standard configuration,
the instrument must be connected to a PC via the USB infrared cable
(order no. 83 17 409). The configuration work is carried out using the
PC software Dräger CC
-
Vision. The PC software Dräger CC Vision
can be downloaded from the following web address free of charge:
www.draeger.com/software.
Change configuration: see Technical Manual.
Standard instrument configuration:
WARNING
CSA requirement: Measured values over the full scale value
may indicate an explosive atmosphere.
Only applies for class II certification. CSA standard C22.2 no.
152 does not contain any requirements for class II hazard
areas. Therefore this device was not tested for class II by the
CSA. The sensor may become blocked and not measure the
gas correctly or not warn the user that gas measurement is
not possible.
WARNING
CSA requirement: The sensitivity must be tested every day
before first use using a known concentration of the gas being
measured corresponding to 25 to 50 % of the full concentration
value. The accuracy must be 0 to +20 % of the daily value. The
accuracy may be corrected via calibration.
NOTICE
CSA requirement: Only the combustible gas detection portion
of this instrument has been assessed for performance.
The instrument has not been classified by the CSA for use in
mines.
!
!
i
i
NOTICE
Only trained persons are permitted to carry out modifications
to the instrument configuration.
Dräger X-am
®
2500
1
Bump test mode
2
Extended bump test
Fresh air calibration
2
ON
Operating signal
2 3
ON
Capture range
8
ON
Switch off
2
allowed
LEL factor
2
(ch
4
)
4.4 (vol. %)
(4.4 Vol. % corresponds to 100 %LEL)
STEL
2 4
5
(short-term average)
STEL function - disabled
Average value duration = 15 minutes
TWA
2 5 6
(shift average)
TWA function - disabled
Average value duration = 8 hours
Alarm A1
7
can be acknowledged, non-latching, pre-alarm,
rising flank
i
i
Summary of Contents for X-AM 2500
Page 2: ......
Page 230: ...230 Dr ger X am 2500 1 245 5 Dr ger Dr ger Dr ger Dr ger 1 1 www draeger com ifu i i i i...
Page 237: ...Dr ger X am 2500 237 CatEx 125 PR Dr gerSensor CatEx 125 PR Gas...
Page 238: ...238 Dr ger X am 2500 4 6 4 6 1 A1 A1 O2 A1 A1 A1 A2 OK 4 6 2 A2 A2 O2 A1 A2 OK CatEx CatEx i i...
Page 247: ......