9
Installing the transmitter
Installing the transmitter in areas subject to explosion hazards of zone 0
or zone 1
Install a safety barrier with the appropriate explosion protection approval
(category 1, 2 or Div. 1) between the transmitter and the control unit.
— Only safety barriers with the following characteristics may be used: U
o
(V
oc
)
≤
30 V, I
o
(Isc)
≤
0.3 A, P
o
≤
700 mW.
— Take care that the maximum permissible capacitance and inductance of connec-
tions to the safety barrier are not exceeded, also taking the cable into account.
The safety-related input parameters of the transmitter are: C
i
= 0 nF, Li = 50 µH.
Transmitter supply units
(without HART-communication between Ex/Non-Ex area)
The following safety barriers are provided as examples only. Selected barriers must
be acceptable to the authority having jurisdiction and comply with the assigned P3S
entity parameters also taking the cable into account.
— Connect shielding to earth point and/or 0 V (Ex i).
Installing the transmitters in explosion-hazard areas of zone 2, or in
areas not subject to explosion hazards
— Use only supply units or a safety barrier classified as device category 3.
— Only supply units or safety barriers with the following characteristics may be
used: U
O
(V
OC
)
≤
30 V, I
O
(I
SC
)
≤
0.3 A, P
O
≤
700 mW.
— Take care that the maximum permissible capacitance and inductance of connec-
tions to the supply unit are not exceeded, also taking the cable into account.
The safety-related input parameters of the transmitter are:
C
i
= 0 nF, Li = 50 µH.
Manufacturer
Type
suitable for
Line (Loop)
MTL
MTL 5041
Zone 0, Div. 1
≤
190
Ω
Pepperl & Fuchs
KFD2–STC4–Ex1
Zone 0, Div. 1
≤
140
Ω
KFD2–STC1–Ex1
Zone 0, Div. 1
≤
140
Ω
CAUTION
The category 1 marking has to be cut out from the rating-plate label. Once the unit
has been used after installation in the above manner, it may never be installed in
explosion-hazard areas of zone 0 or zone 1 (device category 1 or 2).
Explosion hazard!
00433020_1_en.eps
Non-explosion-hazard area
Explosion-hazard area, zone 0, 1 or Div. 1
Safety barrier
Control unit
Ex i
PA
Earth point
4 ... 20 mA
0 V
+
–
+
4 to 20 mA
+24 V
Summary of Contents for Polytron 3000
Page 25: ...25 ATEX approval ATEX approval ...
Page 26: ...ATEX approval 26 ...
Page 27: ...27 ATEX approval ...
Page 28: ...ATEX approval 28 ...
Page 29: ...29 ATEX approval ...
Page 30: ...ATEX approval 30 ...
Page 31: ...31 ATEX approval ...
Page 32: ...ATEX approval 32 ...
Page 33: ...33 ATEX approval ...
Page 34: ...ATEX approval 34 ...
Page 35: ...35 ATEX approval ...
Page 36: ...ATEX approval 36 ...
Page 37: ...37 ATEX approval ...
Page 38: ...ATEX approval 38 ...
Page 39: ...39 ATEX approval ...
Page 40: ...ATEX approval 40 ...
Page 41: ...41 ATEX approval ...
Page 42: ...ATEX approval 42 ...
Page 43: ...43 ATEX approval ...
Page 44: ...ATEX approval 44 ...
Page 45: ...45 ATEX approval ...
Page 46: ...ATEX approval 46 ...
Page 47: ...47 IECEx approval IECEx approval ...
Page 48: ...IECEx approval 48 ...
Page 49: ...49 IECEx approval ...
Page 50: ...IECEx approval 50 ...
Page 51: ...51 UL approval UL approval ...
Page 52: ...UL approval 52 ...
Page 62: ...Declaration of Conformity 62 Declaration of Conformity ...
Page 63: ...63 Declaration of Conformity ...
Page 64: ...64 ...
Page 65: ...65 Drilling templates Dräger docking station 66 4 mm 02723758_1 eps ...
Page 66: ......
Page 67: ......