Maintenance
80
Instructions for use
Polaris 600
Acceptance and handover
Polaris 600 light
Upon conclusion of the installation or maintenance
work, the system must be tested and accepted by
specialized service personnel.
Commissioning DrägerService is recommended.
These tests determine
– whether the safety requirements necessary to
protect the patient and personnel have been
met, and
– whether the entire functional range of the
systems is met.
The inspection results must be documented in
writing.
After acceptance, the operational system is
handed over to its owner along with the associated
documentation. A written record of the handover is
filed for reference.
The users are instructed on the use of the system.
Polaris 600 remote control for MedView camera
It may be necessary to change the room address if
MedView cameras are operated by remote control
in adjacent operating rooms which are only
separated by a pane of glass.
The room address of the remote control and of the
remote control interface can only be changed by
DrägerService.