Instructions for use
Polaris 600
65
Troubleshooting
Wireless video transmission
Fault
Cause
Remedy
Video image is
not displayed
MedView camera is carrying out an initialization after
standby mode.
No remedial action nec-
essary. This procedure
takes up to 30 seconds
Connection between the radio transmitter and the radio
receiver is interrupted
Switch the MedView
camera to standby mode.
Switch the MedView
camera on again after a
few seconds
Connection between the radio transmitter and the radio
receiver is interrupted
Reestablish the connec-
tion:
1
Push up the locking
plate on the radio
receiver housing (dis-
play mount).
2
Set the on/off switch
on the radio receiver
to
0
.
3
Wait for approx. 5
seconds.
4
Set the on/off switch
on the radio receiver
to
I
.
The connection will
be reestablished.
This procedure takes
up to 30 seconds
Video image is
not displayed and
cannot be rees-
tablished
Unknown
Notify DrägerService