Instructions for use
Polaris 600
47
Operation
Polaris 200 (A)
Polaris 200 (B)
Polaris 200 (C)
1740 W/m
2
Acceptable
Polaris 600 (A)
Polaris 600 (B)
Polaris 200 (C)
1780 W/m
2
Acceptable
Polaris 600 (A)
Polaris 200 (B)
Polaris 200 (C)
1720 W/m
2
Acceptable
Polaris 600 (A)
Polaris 600 (B)
Polaris 100 (C)
1660 W/m
2
Acceptable
Polaris 600 (A)
Polaris 100 (B)
Polaris 100 (C)
1520 W/m
2
Acceptable
Polaris 600 (A)
Polaris 200 (B)
Polaris 100 (C)
1620 W/m
2
Acceptable
1) When all lights in the light combination are set to maximum illuminance
2) Maximum permissible total irradiance according to IEC 60601-2-41 corresponds to 1000 W/m
2
Light combina-
tion
Maximum
total irradi-
ance
1)
[W/m
2
]
Recommended adjustment of the total irradi-
ance
2)
by reducing the illuminance
Thermal radiation in
the wound site after
adjustment of the
total irradiance
(A)
(B)
(C)
(C)
(A)
(B)
(C)
(A)
(B)
(C)
(A)
(B)
(C)
(A)
(B)
(C)
(A)
(B)