12.4 Sensor Specifications and Instrument Settings for other Gases
NH
3
SO
2
PH
3
HCN
HCN PC
NO
NO
2
Measuring Range
0 ... 300 ppm 0 ... 100 ppm
0 ... 20 ppm
0 ... 50 ppm
0 ... 50 ppm
0 ... 200 ppm
0 ... 50 ppm
Calibration concentration
50 ppm in N
2
10 ppm in N
2
0.5 ppm in N
2
10 ppm in N
2
15 ppm in N
2
50 ppm in N
2
10 ppm in N
2
Temperature range, operation
–30 ... 50
o
C
–22 ... 122
o
F
–30 ... 50
o
C
–22 ... 122
o
F
–20 ... 50
o
C
–4 ... 122
o
F
–20 ... 50
o
C
–4 ... 122
o
F
-20 ... 50 °C
–4 ... 122
o
F
–40 ... 50
o
C
–40 ... 122
o
F
–30 ... 50
o
C
–22 ... 122
o
F
Alarm Threshold A1
2)
50 ppm
1 ppm
0.1 ppm
10 ppm
5 ppm
25 ppm
5 ppm
acknowledgeable
Yes
Yes
Yes
Yes
Yes
Yes
Yes
latching
No
No
No
No
No
No
No
Alarm Threshold A2
2)
100 ppm
2 ppm
0.2 ppm
20 ppm
10 ppm
50 ppm
10 ppm
acknowledgeable
No
No
No
No
Yes
No
No
latching
Yes
Yes
Yes
Yes
No
Yes
Yes
TWA Threshold A1
2)
50 ppm
1 ppm
0.1 ppm
10 ppm
5 ppm
25 ppm
5 ppm
STEL Threshold A2
2)
50 ppm
1 ppm
0.1 ppm
40 ppm
20 ppm
50 ppm
5 ppm
No. of STEL periods
4
4
4
4
4
4
4
Average STEL duration
15 minutes
15 minutes
15 minutes
15 minutes
15 minutes
15 minutes
15 minutes
Warm up time
12 hours
15 minutes
15 minutes
15 minutes
15 minutes
12 hours
15 minutes
Reproducibility
Zero point:
± 4 ppm
± 0.1 ppm
± 0.02 ppm
± 0.5 ppm
± 3 ppm
±0,3 ppm
± 0.2 ppm
Sensitivity: [% of measured value]
± 3
± 2
± 2
± 5
± 5
±3
± 2
Drift (20
o
C)
Zero point:
± 5 ppm/a
± 1 ppm/a
± 0.05 ppm/a
± 2 ppm/a
± 3 ppm/a
±0,3 ppm/a
± 1 ppm/a
Sensitivity: [% of meas. value/month]
± 2
± 2
± 2
± 5
± 2
±2
± 2
Sensor Order Number
1)
68 10 888
68 10 885
68 10 886
68 10 887
68 13 165
68 11 545
68 10 884
Sensor Data Sheet Order Number
90 23 922
90 23 919
90 23 920
90 23 921
90 33 428
90 33 091
90 23 918
Please be aware of possible sensor cross sensitivities (see Sensor Data Sheet).
1) Please be aware of limited lifetime of sensors. Excessive storage reduces operating time of sensors.
Temperature range of storage is 0 to 35
o
C (32 to 95
o
F)
2) Please be aware of special settings by customer requirements.
3) For ethylene oxide only.