Life of pH probe is approximately 18 months, and may reduce accuracy after long exposure to salt water.  
Please do not allow pH probe to dry completely for 12 hours or more, it could result in permanent damage 
of the probe.

Use of non-approved KHG reagent solution is strictly prohibited.  The factory supplied KH reagent is 
a special mixture of reagents.  We carefully select reputable reagent chemical suppliers with quality 
certifications.  Non-KHG approved reagents may damage the patent pending interference resistant rare 
earth material coating in the Reaction Chamber, and may result in permanent damage to the accuracy.

Factory supplied KHG reagent solution are concentrated.  To achieve the best accuracy, users must proper 
dilute it.  You should use 1500 grams of RODI water for dilution.  Please note the unit of RODI water is 
grams, not mililiter.  Please use a electronic scale to weigh 1500 grams of water at room temperature, and 
keep the error to within 0.1 grams.  Each new dilution of the KHG reagents is enough for approximately 2 
month, based on 4 hours test interval and about 4 grams of reagent consumed per test.

To best protect your reef tank inhabitants, we did not add preservatives to the factory supplied KHG 
reagents.  However, the lack of preservatives may result in black color bacteria substances growing in the 
tubing.  From the test result of using KHG in the last 4 years, these substances do not cause inaccuracy 

to the tests, and do not have negative influences to the reef tank inhabitants. If accidental diluted reagent 

solution exposure to your eyes or skin, please do not panic and simply rinse with large quantities of fresh 
tap water.  Please keep KHG reagent solutions away from children at all times.

KHG currently is not WIFI capable due to various country regulations.  You must use RJ45 network cables 
to physically connect KHG to your LAN.  There is a 12V power jack already provided with KHG that can be 
used in conjunction with RJ45 WIFI Bridge.  If user wish to make KHG WIFI capable, please contact with 
your seller or authorized dealer.  KHG parent company does not offer WIFI related support for the users.

KHG does not support fixed IP and must be attached to the network that uses DHCP.  For any reasons if 
KHG fails to obtain IP from the DHCP server, KHG will fix itself at default settings :

IP : 192.168.1.120  
Submask : 255.255.255.0 
Gateway : 192.168.1.1 
Port : 8090
Please note the above default setting cannot be changed.

If your mobile devices are in this IP range, you can connect to the KHG by entering below in your browser :
http://192.168.1.120:8090 

4.

5.

6.

7.

8.

9.

User Manual

Summary of Contents for KHG

Page 1: ...WWW KHGUARDIAN CO M V8 0C Profashional Edition 2017 01 09...

Page 2: ......

Page 3: ......

Page 4: ...ction Main Display KHG System Setup The Meaning of Log Field Error Message and Troubleshooting Additional User Instructions The KHG exterior features Main Control of KHG P1 P5 P1 P5 P7 P10 P10 P12 P1...

Page 5: ...bicarbonate solution should be added to the aquarium with the reference of the desired dKH target value aquarium total water volume and current testing value Then the controller dose in the solution...

Page 6: ...e sure the placement of KHG is level on the X Y and Z axis Complete levelling is crucial as to keep the sodium bicarbonate solution from dripping on the optical and electronical components before ente...

Page 7: ...eagent solution bottle Please insert the pH probe into pH 7 0 calibration solution of at least 1 5 cm deep Press and hold number 2 and number 4 buttons simultaneously for 3 second until you hear a bee...

Page 8: ...many times as out of the inlet tube Please click on the KS De Gas on your browser to purge air from your KHG reagent solution tube Please look into the observation hole Figure 6 until you see reagent...

Page 9: ...t be performed every time you intend to move the KHG It is crucial to purge reaction chamber to prevent solutions damaging the optical lens and other electronic components Failure to purge chamber cou...

Page 10: ...not use unless being instructed by a qualified KHG support staff ENG Mode OFF Advanced function only Do not use unless being instructed by a qualified KHG support staff Test KH doser Press to test if...

Page 11: ...between each KHG test Unit is minutes We recommend setting the intervals to 180 minutes You can shorten the testing interval time but that will also use up KH reagent solution faster No_SW_ERR This is...

Page 12: ...ntroller KHG not only monitors your tank alkalinity but with KH auto correction KHG will also actively maintain your desired dKH level Regardless if you normally maintain your tank alkalinity by kalkw...

Page 13: ...Save all settings above to the memory KHG will force reboot after new settings have been saved New PW Please enter new password Password allows maximum 8 characters include spaces numbers letters or...

Page 14: ...e Please gently open the Reaction Chamber and make sure the white water level sensor is freely movable up and down Sensor may be malfunctioned Please contact your dealer eg 10 07 15 07 49 W 131 0 AK 3...

Page 15: ...ensor may be malfunctioned Please contact your dealer KH ERR KH value high error A B C D E F 4 This error message indicates the tested value is 1 5dKH larger than the KH auto correction setting System...

Page 16: ...bration procedure If all settings are correct please contact your dealer Page12 4 Additional User Instructions 1 2 3 Upon initial assembly and installation please make sure to perform 5 five dKH readi...

Page 17: ...the factory supplied KHG reagents However the lack of preservatives may result in black color bacteria substances growing in the tubing From the test result of using KHG in the last 4 years these sub...

Page 18: ...mixing 65 grams of sodium bicarbonate powder per one liter of water Please keep the error tolerance to within 5 We would strongly recommend to keep KHG in a protected container away from saltwater sp...

Page 19: ...Press and hold buttons 2 4 for 3 seconds To KS_DS_Gas Press and hold buttons 1 2 for 3 seconds To SW_Tube De_Gas Press and hold buttons 1 3 for 3 seconds To Purge Chamber Press and hold buttons 3 4 fo...

Page 20: ...Contact SALES KHGUARDIAN COM SUPPORT KHGUARDIAN COM...

Reviews: