Total volume (L) :
This is to set the total water volume of your tank in Liters. KH auto correction will use this data to calculate
how much sodium bicarbonate solution are needed after each testing. (Feature not available fo M model.)
SYS_Sound :
Set as “1”, system sound will be ON. Any other number will deactivate system sound. Canceling system
sound will NOT silence alarm.
PORT :
This is to be used only as instructed by your network administrator. Default value is 8090.
Timezone :
This is to set time zone (UTC) at your location.
NTP Server :
This is only to be used if instructed by your network administrator.
MAC :
MAC Address of this KHG and cannot be changed.
S/N :
Serial Number of this KHG and cannot be changed.
SAVE :
Save all settings above to the memory. KHG will force reboot after new settings have been saved.
New PW :
Please enter new password. Password allows maximum 8 characters include spaces, numbers, letters or
mixed. No special character allow. Case sensitive.
Confirm PW :
Please re-enter new password. Press “Change Password” after to complete and store your change. No
reboot necessary.
Clear SD:
Press “DO” to initiate SD card memory purge. KHG has log feature for user analysis. However, the record
log takes up memory on the SD card, and we recommend performing the SD card memory purge every 3
months. Low available SD card memory will slow down the display refresh speed. Every time you do a “Clear
SD”, please unplug KHG from power outlet for 3 seconds and reconnect the power cord to the electrical
outlet. This will reboot the system to re-enable the logging function.